1/2

品牌: R&D Systems









产品介绍
产品信息
产品详情
Product Summary
The Parameter Estradiol Immunoassay is a 3.5 hour competitive enzyme immunoassay designed to measure Estradiol in cell culture supernates, serum, and plasma.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of a known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of kit components
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 153 | 744 | 1419 | 164 | 765 | 1533 |
Standard Deviation | 13.6 | 26.8 | 76.6 | 9.54 | 61.5 | 116 |
CV% | 8.9 | 3.6 | 5.4 | 5.8 | 8 | 7.6 |
Recovery
The recovery of Estradiol spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 94 | 81-102 |
EDTA Plasma (n=4) | 103 | 90-113 |
Heparin Plasma (n=4) | 102 | 88-117 |
Serum (n=4) | 96 | 87-108 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Estradiol were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data
Assay Procedure
Refer to the product Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate.
- Prepare all reagents, standard dilutions, and samples as directed in the product insert.
- Remove excess microplate strips from the plate frame, return them to the foil pouch containing the desiccant pack, and reseal.
- Add 100 µL of Estradiol Primary Antibody Solution to each well (excluding the NSB wells). Cover with a plate sealer, and incubate at room temperature for 1 hour on a horizontal orbital microplate shaker.
- Aspirate each well and wash, repeating the process 3 times for a total of 4 washes.
- Add 100 µL of Standard, control, or sample to the appropriate wells.
- Add 100 µL of Calibrator Diluent RD5-62 to the zero standard (B0) wells and NSB wells.
- Add 50 µL of Estradiol Conjugate to all wells. Cover with a new plate sealer, and incubate at room temperature for 2 hours on the shaker.
- Aspirate and wash 4 times.
- Add 200 µL Substrate Solution to each well. Incubate at room temperature for 30 minutes on the benchtop. PROTECT FROM LIGHT.
- Add 100 µL of Stop Solution to each well. Read at 450 nm within 30 minutes. Set wavelength correction to 540 nm or 570 nm.
100 µL Estradiol Primary Antibody Solution


100 µL Standard, Control, or Sample


100 µL Calibrator Diluent RD5-62


50 µL Estradiol Conjugate


200 µL Substrate Solution


100 µL Stop Solution


Estradiol Parameter Assay Kit Summary
Format
96-well strip plate
Assay Length
3.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL)
Sensitivity
12.1 pg/mL
Assay Range
12.3 - 3,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Measures Estradiol levels in various samples. This assay cross-reacts 0.07% with 17 alpha -ethynylestradiol, 0.03% with 17 beta -estradiol 3-Benzoate, 0.86% with Estriol, 0.26% with Estrone and 0.06% with Progesterone.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity observed with 1 or more species tested.
Interference
Interference observed with 1 or more available related molecules.
背景
研究领域
制备和贮存
保存方式
Preparation and Storage
Shipping
The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.
声明 :本官网所有报价均为常温或者蓝冰运输价格,如有产品需要干冰运输,需另外加收干冰运输费。